|
QUARTZ ANALYTICAL METHOD (P-7)
INFRARED DETERMINATION OF QUARTZ
IN RESPIRABLE COAL MINE DUST
| ANALYTE: |
Quartz |
METHOD NO.: |
P-7 |
| MATRIX: |
Respirable Coal Mine Dust |
RANGE: |
25-250 micrograms of quartz |
| PROCEDURE: |
Infrared |
PRECISION: |
5-10% RSD |
| DATE ISSUED: |
August 3, 1989 |
QUANTIFICATION LIMIT: |
10 µg |
| DATE REVISED: |
November 14, 1994 |
1 PRINCIPLE OF THE METHOD
1.1 Airborne respirable dust samples are collected on
membrane filters (in capsules pre-weighed to one
hundredth of a milligram, or one thousandth of a
milligram) using MSHA/NIOSH approved personal respirable
dust samplers as described in 30 CFR Part 74.
1.2 After collection, the filter capsules are reweighed to
one hundredth of a milligram in order to determine the
net sample mass.
1.3 The sample filters are ashed in a low-temperature radio-frequency asher to destroy the organic matrix (coal dust
and collection filter).
1.4 Ashed samples are deposited onto a 0.64 cm2 circular area
(9 mm diameter) of a DM-450 vinyl/acrylic copolymer
filter. (The use of smaller deposit areas may increase
the sensitivity of the analysis. Redeposited samples to
be analyzed for quartz must be of the same deposit
diameter as calibration samples.)
1.5 The redeposited, ashed dust samples are scanned by
infrared spectrometry between frequencies of 1,000 and
700 cm-1 to determine the quartz and kaolinite content.
1.6 The mass of quartz in the deposit is determined (after
correcting for the interference by kaolinite) using
calibration data from standard quartz samples and
standard kaolinite samples.
1.7 The percentage of quartz in the samples is calculated
using the quartz mass determined from the analysis and
the and the samples mass of dust.
2 RANGE AND LIMIT OF DETECTION
2.1 The range of this method is 25 to 250 micrograms of
quartz. The mass required is 0.45 milligram of
respirable coal mine dust if filters are preweighed to
0.01 milligram and 0.100 milligram of dust if preweighed
to 0.001 milligram.
2.2 The detection limit of this method (3 time the standard
deviation of the analysis of 10 blank filters) is 3
micrograms quartz.
3 INTERFERENCES
3.1 Cristobalite, tridymite, kaolinite, and amorphous silica
have absorbance peaks at 800 cm-1. Of these, however,
only kaolinite has been detected in coal mine dust.
3.2 Coal mine dust frequently contains kaolinite. Since
kaolinite has an absorption band at the same frequency as
quartz (800 cm-1), its presence causes an overestimation
of the quartz mass. However, kaolinite also has an
infrared absorbance band at 917 cm-1. (The kaolinite
absorbance at 917 cm-1 is approximately eight times that
at 800 cm-1.) The kaolinite content in the sample is
determined from the 917 cm-1 band and the kaolinite
contribution to the absorbance at 800 cm-1 is determined
from the absorbance of kaolinite at 917 cm-1 using the
calibration data. The quartz absorbance is the remainder
of the absorbance at 800 cm-1 after subtracting the
kaolinite contribution from the total absorbance.
4 PRECISION AND ACCURACY
4.1 The error of the low-temperature asher, infrared method
for the determination of quartz mass is between 5 and 10
percent relative standard deviation (RSD in the 100 to
500 microgram quartz range.(1)
4.2 The accuracy of the method may depend on the particle
size distribution and purity of standard materials used
in preparation of instrument calibration.
5 ADVANTAGES AND DISADVANTAGES OF THE METHOD
5.1 Advantages
5.1.1 The quartz analysis can be performed on a single
sample having a weight gain as low as
0.100 milligram if the sample is pre- and post-weighed to a thousandth of a milligram, or a
weight gain as low as 0.45 milligram if the sample
is weighed to a hundredth of a milligram.
5.1.2 The detection limit is lower than either the X-ray
diffraction method (Anderson 1983) or the Talvitie
method (Talvitie 1951, 1958).
5.1.3 Samples prepared for analysis can also be analyzed
using X-ray diffraction without any additional
preparation.
5.1.4 The analytical equipment is considerably less
expensive than that required for X-ray
diffraction.
5.2 Disadvantages
The method requires the low-temperature ashing and
redeposition of all samples prior to analysis.
6 APPARATUS
6.1 Fourier transform infrared spectrophotometer; PE Model
1750, 1725x or equivalent with optional sample shuttle
accessory.
6.2 Polystyrene film standard (0.05 millimeter thickness); PE No. 186-2082 or equivalent.
6.3 Low-temperature radio-frequency (RF) asher and vacuum
pump; LFE Model LTA-504 or equivalent.
6.4 Oxygen tank and two-stage regulator for low temperature
asher, second stage of regulator capable of being set at
2 to 10 psi.
6.5 Microbalance capable of weighing to 0.001 or
0.01 milligrams; Mettler M-3 or equivalent.
6.6 Filters for sample redeposit; Gelman Instrument Company,
DM-450, vinyl/acrylic copolymer membrane, 47 millimeter
diameter, 0.45 micrometer pore size.
6.7 Filters for supporting collection filter during
redepositing; Gelman Type A/E glass fiber filters,
25 millimeter diameter.
6.8 Filtration apparatus consisting of: 3-place manifold,
fritted supports and clamps, rubber stoppers, a 4,000 ml
side-arm filtering flask and vacuum tubing. (See Figure 1
for illustration of this apparatus.)
6.9 Vacuum source (water aspirator or vacuum pump) with trap,
(2,000 ml side-arm flask). (See Figure 1 for
illustration of this apparatus.)
6.10 Small, stainless steel forceps for opening filter
cassettes and for handling filters.
6.11 Petrislides for 47 millimeter diameter filters; Millipore
No. PD15-047-00 or equivalent.
6.12 Ultrasonic bath, 200 watt input.
6.13 Wash bottles, polyethylene, 250 milliliter.
6.14 Beakers, Pyrex, 50 ml, one per sample to be analyzed
(Beakers are numbered in sequence and grouped in lettered
sets to prevent mixing of samples, i.e., A1, A2,...A28;
B1, B2,...B28; etc.)
6.15 Diamond Marking Pencil (used to mark numbers on beakers
in Section 6.14).
6.16 Aluminum filtering funnels (These are specially
fabricated funnels, similar to the Millipore No.
XX10-025-40, and are shown on Figure 2. All funnels used
for calibration and analysis must have the same bore
diameter, such as the 9 millimeters illustrated.)
6.17 Desiccator for storing the quartz and kaolinite standard
materials.
6.18 Sample holders for infrared instrument (These are
specially-made steel plates with a center hole having a
diameter the same as, or slightly smaller than, that of
the sample deposit (9 mm). Small ring magnets are used to
hold the filters in position on the sample holder.)
6.19 Slide warmer (for drying filters after filtering).
6.20 Volumetric pipettes, Class A, Pyrex-type or borosilicate
glass, in the following sizes: 0.5, 1, 2, 3, 4, 5, 6, 7,
8, 9, 10, 15, 20, 25, and 30 milliliters (for preparing
calibration standards and QA samples).
6.21 Pipette filler, rubber or neoprene.
6.22 Flasks, volumetric, Pyrex-type, with stoppers, 500 and
1,000 milliliter sizes.
6.23 Spatula, micro, stainless steel, (for weighing out quartz
and kaolinite standard materials).
6.24 Weighing boats, aluminum, 12 millimeter; Cahn Instruments
No. 1187.
7 MATERIALS
7.1 Isopropanol (IPA), technical and ACS reagent grade.
7.2 Quartz for calibration samples; Standard Reference
Material(SRM)1878, "Respirable Alpha Quartz," U.S.
Department of Commerce, National Institute for Standards
and Technology, Washington, D.C. or -5 micrometer
MINUSIL, Pennsylvania Glass and Sand Company, Berkeley
Springs, West Virginia.
7.3 Kaolinite for calibration samples; Georgia Kaolin,
Hydrite UF or equivalent.
8 PROCEDURE
8.1 Assemble the numbered beakers, the respirable coal mine
dust samples to be analyzed, and the sample handling data
sheet that is obtained from the computer. (See Sample
Handling Procedure-Quartz Lab, April 22, 1993.)
8.2 Using the forceps, remove the collection filter and
stainless steel wheel from the cassette capsule and place
them, first the wheel and then the filter with the dust
side down, into the bottom of a numbered beaker.
8.3 Put the beakers containing the samples into the low-temperature asher. With each run of filters ashed (up to
25), also include three quality assurance (QA) samples.
(See Internal Quality Assessment Program for Quartz
Analysis, October 1992 for an explanation of the QA
samples.)
8.4 Ash samples until the filters are completely ashed,
approximately 1 hour, using 300 watts RF power and an
oxygen flow rate of approximately 200-300 cc/minute.
(See Standard Operating Procedures Low-Temperature
Ashers.) Carefully remove the glass beakers from the
low-temperature asher. As each beaker is removed, gently
rinse the sides with about 10 ml of technical grade IPA
using a wash bottle. Do not squirt the IPA directly at
the bottom of the beaker as this will blow the ash out of
the beaker.
8.5 Set-up the filtering apparatus (Section 6.8) as
illustrated in Figure 1. Place two Type A/E glass fiber
filters on top of each fritted support.
8.6 For each sample to be filtered, cut a 47 mm DM-450 filter
into quarters. Place one quarter, glossy side down, on
top of the Type A/E glass fiber filter.
8.7 Position the filter funnel on top of the DM-450 filter
and clamp securely in place.
8.8 Add several milliliters of technical grade isopropanol
(IPA) to the funnel with a wash bottle. Apply the vacuum
and check for leakage at the base of the funnel. If
leakage occurs, unclamp and reposition funnel, then
recheck for leakage. Turn off vacuum.
8.9 Place the beakers containing the dust/alcohol slurry into
the ultrasonic bath for 1 to 2 minutes to disperse the
dust.
8.10 Remove the beakers from the bath, wipe any excess water
from the outside, then carefully pour the slurry into the
filter funnel, being careful not to lose any drops of
alcohol. Apply the vacuum.
8.11 Rinse the inside of the beaker with IPA from the wash
bottle and add the rinsing to the funnel. Repeat this
rinsing a second time.
8.12 When the depth of the liquid in the funnel reaches about
2 centimeters above the filter, gently rinse the inside
of the funnel with IPA and filter until all alcohol is
removed. When the alcohol is below the 'lip' on the
funnel, carefully rinse the 'lip' on the inside of the
funnel so as not to disturb the ash deposit.
8.13 When filtration is complete, remove the clamp and lift
off funnel while taking care not to scrape off or disturb
the deposit on the DM-450 filter. Release the vacuum.
8.14 Remove each filter from the filter base. Number the
corner of the filter with the beaker number and place the
filter in a petrislide which is numbered and lettered the
same as the beaker so that the samples are not mixed up.
8.15 Prepare a filter blank by filtering IPA through a DM-450
filter quarter as in Sections 8.5 through 8.8. Place the
blank filter in a petrislide marked "blank".
8.16 Set the petrislides containing the filters on the slide
warmer with the lids partially ajar to permit
evaporation. Several holes, approximately 1 mm in
diameter drilled through the top of the petrislides
ensure thorough evaporation of the alcohol. Allow the
filters to dry for about 30 minutes. Remove the
petrislides from the slide warmer and close the lids.
Note: Drying samples in this manner is necessary due to
the presence of an interfering peak, created from the
combination of the alcohol, polystyrene (petrislide) and
the heat from the hotplate, which causes an error in the
baseline position and the quartz peak height measurement.
8.17 Set the instrument parameters on the FTIR as follows:
| Resolution: |
8 cm-1 |
| Acquire mode: |
Interleaved (requires sample shuttle accessory) |
| Apodization: |
Normal |
| Number of scans: |
1 |
Note: When using the FTIR, the setting of these
parameters and the following sections, 8.18
through 8.20 are performed through computer
programs.
8.18 Prior to scanning any set of samples or standards, check
the operation of the infrared spectrophotometer using a
polystyrene film standard. Scan the polystyrene standard
between 1,000 and 850 cm-1. An absorption band should
occur at 908 cm-1. The intensity (peak height) of the
absorption band from a baseline drawn between the
shoulders on each side of the peak should be within the
limits established for the film on the particular
instrument being used.
Note: this intensity and its limits vary from instrument
to instrument and between one standard film and another,
and must be established by direct observation over a
period of time. If either or both of these conditions
are not met, then it may indicate that there is a problem
with the spectrophotometer. No samples should be
analyzed until the problem is brought to the supervisor's
attention and until corrective action is taken as
dictated by the operator's manual or repair personnel.
The polystyrene film is used as a standard and must be
cared for accordingly. Keep polystyrene films in
suitable containers to prevent them from becoming dirty
or scratched.
8.19 Scan all samples between frequencies of 1,000 and 700 cm-1, beginning with the three QA samples. If quartz
is present in a sample, a doublet will be present in the
scanned range, the higher frequency peak at 800 cm-1.
Kaolinite in the sample is recognizable as an absorption
band at 917 cm-1, usually on the shoulder of a broad,
more intense band. (If both quartz and kaolinite are
present in a sample, the kaolinite band in the 800 cm-1
region is obscured by the quartz doublet.) The infrared
scan for a typical field sample containing both quartz
and kaolinite is shown in Figure 3.
8.19.1 Remove one filter at a time, so as not to mix
samples, from its petrislide holder and place it
onto an infrared sample holder (Section 6.18).
Use a light source to center the deposit over the
hole in the holder, then secure the filter with a
ring magnet.
8.19.2 Insert the holder with the filter sample into the
sample position of the sample shuttle.
8.19.3 Place the dried blank filter onto a holder,
secure with a ring magnet, and insert into the
reference position of the sample shuttle.
8.20 Determine the results for all samples, beginning with the
three QA samples. The QA samples are a test of the total
system operation: Low-temperature ashing, the filtering
process, and infrared scanning. If the QA sample results
are not within established limits, notify the laboratory
supervisor. If the QA sample results are within the QA
program limits, then determine the quartz content of all
samples ashed. (See Internal Quality Assessment Program
for Quartz Analysis, October 20, 1992 for an explanation
of the QA samples.)
8.20.1 When not automatically performed through computer
programs, determine the peak heights for the
absorption bands for each sample according to the
following steps.
8.20.1.1 Draw a baseline for the quartz/kaolinite
absorbance band between 815 and 770 cm-1.
8.20.1.2 Draw a baseline for the kaolinite
absorbance band between 950 and 895 cm-1 (see Figure 3).
8.20.1.3 For each peak, draw a straight line from
the vertex of the peak at either 917 or
800 cm-1 down through the baseline drawn
in Sections 8.20.1.1 and 8.20.1.2.
8.20.1.4 Determine the height of each peak in
absorbance units, from the peak vertices
to the point where the vertical lines
intersect the baselines.
8.20.2 Calculate the corrected absorbance for quartz at
800 cm-1:
8.20.2.1 Determine the kaolinite contribution to
the quartz/kaolinite peak at 800 cm-1
from the calibration data (Section 9):
| kaolinite abs.@ 800 cm-1 = |
| kaolinite abs.@ 917 cm-1 X mean abs.@800 |
| abs.@917 |
8.20.2.2 Subtract the kaolinite absorbance
contribution to the 800 cm-1
quartz/kaolinite peak from the total
absorbance:
corrected abs. quartz =
abs. quartz + kaolinite @800 cm-1 - abs. kaol. @ 800 cm-1
8.20.3 Calculate the mass of quartz present in the
samples using the calibration information
(Section 9):
| mass of quartz (µg) = corrected absorbance |
| mean abs./microgram |
8.21 To determine the percent quartz, divide the mass of quartz
(micrograms) by the sample mass (converted to micrograms)
and multiply by one hundred. The precision of the percent
determination is dependent on the precision of the sample
mass determination.
9 CALIBRATION
9.1 Kaolinite
9.1.1 Prepare a suspension of kaolinite in IPA,
100 micrograms of kaolinite per milliliter of
alcohol.
9.1.1.1 Weigh out 25.000 ±0.001 milligrams of
dried kaolinite.
9.1.1.2 Quantitatively transfer the kaolinite to
a 500 milliliter volumetric flask,
adjust the temperature of the suspension
to 20°C using either hot or cold water
baths as necessary, and bring to volume
with reagent grade IPA.
9.1.1.3 Disperse the kaolinite by placing the
flask in an ultrasonic bath for
30 minutes.
9.1.1.4 Cool the flask in a cold water bath
until the temperature of the suspension
returns to 20°C.
9.1.2 Prepare kaolinite standards using the same
filtering apparatus described in Sections 6.8,
6.9, and 6.16 and illustrated in Figure 1.
Position two Type A/E glass fiber filters and one
quarter of a DM-450 filter on each fritted
support as described in Sections 8.5 and 8.6.
Position and clamp the filtering funnel in place
and visually check for leaks as indicated in
Sections 8.7 and 8.8 using reagent grade IPA.
9.1.3 Prepare at least five DM-450 filters containing
the following quantities of the kaolinite
standard material: 50, 100, 300, and 500 and 750
micrograms. These are obtained by pipetting 1,
2, 6, 10, and 15 milliliters, respectively, of
the standard suspension onto the filters.
9.1.3.1 Vigorously shake the flask with the
kaolinite suspension about 15 times.
9.1.3.2 Withdraw an aliquot of the suspension
from the center of the flask with the
appropriate pipette, then allow this to
drain back into the flask to condition
the pipette. Repeat.
9.1.3.3 Draw the liquid up to the mark on the
pipette, adjusting as necessary.
9.1.3.4 Drain the suspension into the filter
funnel. Do not blow out the pipette.
9.1.3.5 Apply vacuum to the funnel and complete
the preparation of the kaolinite
standard as described in Sections 8.13
to 8.16.
9.1.3.6 Complete preparation of the remaining
kaolinite standards as described above.
Shake the flask three or four times
before each aliquot is withdrawn.
9.1.4 When dry, scan the standards on the infrared
spectrophotometer between the frequencies of
1,000 and 700 cm-1, following Sections 8.17 to
8.19.
9.1.5 Calculate the absorbance of kaolinite at 917 cm-1
and at 800 cm-1, as described in Section 8.20.1.1
through 8.20.1.4, for each of the kaolinite
standards.
9.1.6 Determine the ratio of the net absorbance at
917 cm-1 to the net absorbance at 800 cm-1 for
each sample.
9.1.7 Determine the mean ratio for the five samples.
This ratio is used to determine the kaolinite
contribution to the 800 cm-1 quartz peak.
9.1.8 Calculate the absorbance per microgram of
kaolinite for each of the standards.
| ex.: |
| absorbance per microgram = absorbance @ 917 cm-1 |
| mass, µg |
9.1.9 Calculate the mean absorbance per microgram for
the five standards. This value is used to
determine the mass of kaolinite in a sample.
9.2 Quartz
9.2.1 Prepare a suspension of -5 micrometer MINUSIL
quartz in isopropyl IPA, 10 micrograms of MINUSIL
per milliliter of alcohol.
9.2.1.1 Weigh out 10.00 ±0.01 milligrams of
dried -5 micrometer MINUSIL.
9.2.1.2 Quantitatively transfer the MINUSIL to a
1,000 ml volumetric flask, adjust the
temperature of the suspension to 20C
using either hot or cold water baths as
necessary, and bring to volume with
reagent grade IPA.
9.2.1.3 Disperse the MINUSIL by placing the
flask in an ultrasonic bath for 30
minutes.
9.2.1.4 Cool the flask in a cold water bath
until the temperature of the suspension
returns to 20C.
9.2.2 Prepare the quartz standards using the same
filtering apparatus described in Sections 6.8,
6.9, and 6.16, and Figure 1. Position two Type
A/E glass fiber filters and one quarter of a
DM-450 filter on each fritted support as
described in Sections 8.6 and 8.7.
9.2.3 Prepare at least five DM-450 filters containing
the following quantities of the quartz standard
material: 20, 50, 100, 200, and 300 micrograms.
These are obtained by pipetting 2, 5, 10, 20, and
30 milliliters, respectively, of the standard
suspension onto the filters. Complete
preparation of the standards following sections
9.1.3.1 through 9.1.4.
9.2.4 Calculate the absorbance of quartz at 800 cm-1
for each standard, as in Section 8.20.1.1,
8.20.1.3 and 8.20.1.4.
9.2.5 Calculate the absorbance per microgram of quartz
for each of the quartz standards.
9.2.6 Calculate the mean absorbance per microgram for
the five quartz standards. This value is used to
determine the mass of quartz in the sample.
BIBLIOGRAPHY
Anderson, C. C. Collaborative Tests of Two Methods for Determining Free Silica in Airborne Dust. DHHS (NIOSH)
Publication No. 83-124, Contract No. 210-79-0059, February
1983.
Freedman, R. W., S. Z. Toma, and H. W. Lang. On Filter Analysis of Quartz in Respirable Coal Dust by Infrared Absorption and
X-ray Diffraction. Am. Ind. Hyg. Assoc. J., v. 35, 1974,
p. 411.
Talvitie, N. A. Determination of Quartz in Presence of Silicates
Using Phosphoric Acid. Analytical Chemistry, v. 23, No. 4, April 1951, pp. 623-626.
Talvitie, N. A., and F. Hyslop. Colorimetric Determination of Siliceous Atmospheric Contaminants. Am. Ind. Hyg. Assoc. J.,
v. 19, No. 1, February 1958, pp. 54-58.
| About Us | Asbestos | Commercial | Fen - Phen | Silicos | Breast Implants | Location |
 
|